Brand: Analytical Technology IncATI’s Series B12 Wet Gas Detector was developed to meet the need for reliable measurements in wet scrubber outlets for other gas streams that might be saturated with water vapor.
Gas detector systems have been developed for monitoring strong oxidants (chlorine, bromine, chlorine dioxide, or ozone), acid gases (HCl or HF), and hydrogen.
The basic B12 Wet Gas Detector system consists of a loop-powered 2-wire transmitter and a corrosion resistant sensor specifically engineered to operate at 100% RH.
Gas sensors are designed for direct insertion into vent stacks, or may be used in a flow-cell if a sample can be pumped to the sensor.
When remote display and alarm relay functions are required, the B12 wet gas transmitters can be connected to ATI's B14 Gas Alarm Module.
This receiver system will provide direct digital display of gas concentrations, plus 3 alarm set points, a remote alarm reset input, and retransmitted 4-20 mA output.
The B12 wet gas transmitter may also be used separately and tied directly to the user's DCS, PLC, or other equipment.
Either way, the B12 Wet Gas Detector is an excellent solution to a difficult gas monitoring problem.
Features- Minimal maintenance
- Rapid response
- Transmitter LCD display option
- Loop-powered electronics
Specifications
Gas Type
- Oxidant gases: Cl2, Br2, ClO₂, O3
- Acid gases: HF, HCl
- Hydrogen
Range
- 0-10, 0-20, 0-50, or 0-100 ppm for Oxidant or Acid Gases
- 0-2 % or 0-10 % for Hydrogen
Accuracy
- Generally ± 5 % of Value, but Limited by Available Calibration Gas Accuracy
Electronic Repeatability
- ± 1 % of Span
Electronic Lineraity
- ± 0.5 % of Span
Output
- Loop-Powered 4-20 mA, 675 ohms Max. at 24 Vdc
Power
- 12-28 Vdc
Enclosure
- NEMA 4X Polystyrene, LCD Display Optional. Explosion-Proof Optional, Class 1, Div. 1, Groups B, C, & D
Controls
- Zero and Span Potentiometers Mounted on Transmitter Circuit Board
Operating Temperature
- 0° to +50° C
Sensor Materials
- Noryl
Sensor Cable
- 25 ft standard, 100 ft max.
Weight
- 1 lb (0.4 kg)