Yodify Product Library
Add the AP3203 Omnidirectional Hotwire Probe to your store or catalog
Book Your Demo and See How
or create your store
AP3203 Omnidirectional Hotwire Probe
Brand: Delta OHM S.R.L.
1 - 3 of 3 Part Numbers

AP3203
Measuring range: air speed 0.02 ÷ 5 m/s, temperature 0…+80 °C. Probe stem Ø 8 mm, length 230 mm. Cable length 2 metres. Equipped with SICRAM module.
Sensor: NTC 10 kΩ
Resolution: 0.01 m/s
Accuracy: ± (0.05 + 5% measure) m/s
Temperature drift@ 20 °C: 0.06 %/°C
Long term stability: 0.12 °C/year
Connection 8-pole female DIN45326 connector
Protection dimensions Ø=80 mm
In combination with the data loggers HD32.1 and HD32.3TCA, the probe is used for calculating the following indices:
- PMV – Predicted Average Rating
- PPD – Dissatisfied percentage
- TU – Turbulence
- DR – Draught rate
The probe is also used for calculating Mean radiant temperature.

AP3203.2
Measuring range: air speed 0.02 ÷ 5 m/s, temperature 0…+80 °C. Probe stem Ø 8 mm, length 230 mm. Cable length 2 metres. Equipped with SICRAM module.
Sensor: NTC 10 kΩ
Resolution: 0.01 m/s
Accuracy: ± (0.05 + 5% measure) m/s
Temperature drift@ 20 °C: 0.06 %/°C
Long term stability: 0.12 °C/year
Connection 8-pole female DIN45326 connector
Protection dimensions Ø=80 mm
In combination with the data loggers HD32.1 and HD32.3TCA, the probe is used for calculating the following indices:
- PMV – Predicted Average Rating
- PPD – Dissatisfied percentage
- TU – Turbulence
- DR – Draught rate
The probe is also used for calculating Mean radiant temperature.

AP3203F
Measuring range: air speed 0.02 ÷ 5 m/s, temperature -30…+30 °C. Probe stem Ø 8 mm, length 230 mm. Cable length 2 metres. Equipped with SICRAM module.
In combination with the data logger HD32.1, the probe is used for calculating the following indices:
- IREQ – Required insulation
- DLE – Exposure time limit
- RT – Required recovery time
- WCI – Wind Chill Index
- PMV – Predicted Average Rating
- PPD – Dissatisfied percentage
- DR – Draught rate
The probe is also used for calculating Mean radiant temperature.
1 - 3 of 3 Part Numbers